Calibration of an irradiation facility
dc.contributor.author | Marcinko, Rion, author | |
dc.contributor.author | Johnson, Thomas, advisor | |
dc.contributor.author | Brandl, Alexander, committee member | |
dc.contributor.author | Duncan, Colleen, committee member | |
dc.date.accessioned | 2017-06-09T15:41:09Z | |
dc.date.available | 2017-06-09T15:41:09Z | |
dc.date.issued | 2017 | |
dc.description.abstract | The objective of this research is to provide accurate reliable data for use in an irradiation facility and present findings in a manner useful to operators. The intended use of the irradiator and irradiator protocols dictate a need for a well-known administered dose. Dose rates from two independent methods were investigated at multiple distances away from the source. First, the irradiation platforms were examined for "hot spots" using Gafchromic © film to provide a qualitative estimate of the dose distribution. A quantitative assessment of doses was then performed using Fricke dosimetry (a primary standard). Finally, MCNP modeling was used to simulate irradiation at the various measured points and intermediate points in the radiation field. The results were used to provide dose rates within the radiation field for future researchers, and demonstrated how precision and accuracy vary using theoretical (MCNP) and measurement methodology. MCNP simulation matched the Fricke measurements within 10% for the first two positions explored, then diverged upon further movement from the source. | |
dc.format.medium | born digital | |
dc.format.medium | masters theses | |
dc.identifier | Marcinko_colostate_0053N_14080.pdf | |
dc.identifier.uri | http://hdl.handle.net/10217/181356 | |
dc.language | English | |
dc.language.iso | eng | |
dc.publisher | Colorado State University. Libraries | |
dc.relation.ispartof | 2000-2019 | |
dc.rights | Copyright and other restrictions may apply. User is responsible for compliance with all applicable laws. For information about copyright law, please see https://libguides.colostate.edu/copyright. | |
dc.title | Calibration of an irradiation facility | |
dc.type | Text | |
dcterms.rights.dpla | This Item is protected by copyright and/or related rights (https://rightsstatements.org/vocab/InC/1.0/). You are free to use this Item in any way that is permitted by the copyright and related rights legislation that applies to your use. For other uses you need to obtain permission from the rights-holder(s). | |
thesis.degree.discipline | Environmental and Radiological Health Sciences | |
thesis.degree.grantor | Colorado State University | |
thesis.degree.level | Masters | |
thesis.degree.name | Master of Science (M.S.) |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- Marcinko_colostate_0053N_14080.pdf
- Size:
- 682.11 KB
- Format:
- Adobe Portable Document Format