Repository logo
 

Evaluation of the performance of the Gulton Ultrasonic, Gentile Tube and Potter model 6-424 GLMD-5 flowmeters

dc.contributor.authorVideon, Fred, author
dc.contributor.authorSmith, George L., author
dc.contributor.authorColorado State University Research Foundation, publisher
dc.date.accessioned2017-07-18T13:52:04Z
dc.date.available2017-07-18T13:52:04Z
dc.date.issued1958-06
dc.descriptionCER59FV11.
dc.descriptionJune 1958.
dc.descriptionConducted for the Martin Company, Denver Division, Denver, Colorado through Colorado State University Research Foundation, Civil Engineering Section, Fort Collins, Colorado.
dc.description.abstractThis report represents an investigation and evaluation of the performance of several types of flowmeters under simulated prototype conditions. Three types of flowmeters were tested: Gulton Ultrasonic, Gentile Tube, and Potter Model No. 6-424 GLMD-5. The Gulton Ultrasonic Meter on the basis of initial test results was found to be of faulty design, and was, therefore, excluded from further testing. Meter performance for changes in flow and fluid properties was studied. Changes in flow properties included both low and high backpressure intensities, and turbulence induced immediately upstream of the meter by means of a grid. Injection of air into the test line produced a change in the mass density of the fluid - - a fluid property. Meter sensitivity in recording discharge were also made using combinations of peizometric taps located at various positions along the boundary geometry. A comparison is made of the performance of the flowmeters under similar changes in flow and fluid properties.
dc.format.mediumreports
dc.identifier.urihttps://hdl.handle.net/10217/182576
dc.languageEnglish
dc.language.isoeng
dc.publisherColorado State University. Libraries
dc.relationCatalog record number (MMS ID): 991012051339703361
dc.relationTA7.C6 CER 59-11
dc.relation.ispartofCivil Engineering Reports
dc.relation.ispartofCER, 59-11
dc.rightsCopyright and other restrictions may apply. User is responsible for compliance with all applicable laws. For information about copyright law, please see https://libguides.colostate.edu/copyright.
dc.subjectFlow meters -- Testing
dc.subjectFlow meters -- Evaluation
dc.subjectHydraulic engineering
dc.titleEvaluation of the performance of the Gulton Ultrasonic, Gentile Tube and Potter model 6-424 GLMD-5 flowmeters
dc.typeText
dcterms.rights.dplaThis Item is protected by copyright and/or related rights (https://rightsstatements.org/vocab/InC/1.0/). You are free to use this Item in any way that is permitted by the copyright and related rights legislation that applies to your use. For other uses you need to obtain permission from the rights-holder(s).