Repository logo
 

Temperature sensitivity of Eppley broadband radiometers

Date

1988-02

Authors

Smith, W. L., Jr., author
Cox, Stephen K., author
Glover, Vince, author

Journal Title

Journal ISSN

Volume Title

Abstract

Eppley Laboratory Inc. model PIR pyrgeometers and model PSP pyranometers have built in temperature compensation circuitry designed to limit relative errors in the measurement of radiation to + /- 2% for a temperature range of -20 C to +40 C. A procedure developed to verify this specification and to determine the relative sensitivity to temperatures below -20 C is described. Results of this calibration and application to data correction are also presented.

Description

February 1988.
FIRE series no. 5.

Rights Access

Subject

Radiation -- Measurement
Radiometers
Pyranometer
Pyrgeometer

Citation