Smith, W. L., Jr., authorCox, Stephen K., authorGlover, Vince, author2022-05-032022-05-031988-02https://hdl.handle.net/10217/234915February 1988.FIRE series no. 5.Eppley Laboratory Inc. model PIR pyrgeometers and model PSP pyranometers have built in temperature compensation circuitry designed to limit relative errors in the measurement of radiation to + /- 2% for a temperature range of -20 C to +40 C. A procedure developed to verify this specification and to determine the relative sensitivity to temperatures below -20 C is described. Results of this calibration and application to data correction are also presented.reportsengCopyright and other restrictions may apply. User is responsible for compliance with all applicable laws. For information about copyright law, please see https://libguides.colostate.edu/copyright.Radiation -- MeasurementRadiometersPyranometerPyrgeometerTemperature sensitivity of Eppley broadband radiometersText